Research and Education Marginal fit of CAD-CAM monolithic zirconia crowns fabricated by using cone beam computed tomography scans
The Journal of Prosthetic Dentistry
Available online 23 October 2019
Abstract
Statement of problem
Information
regarding the precision of monolithic zirconia crowns fabricated by
using a standard computer-aided design and computer-aided manufacturing
(CAD-CAM) workflow is available. However, information on the effect of a
modified workflow using 3D laboratory scanning and/or cone beam
computed tomography (CBCT) for monolithic zirconia crown fabrication is
lacking.
Purpose
The
purpose of this in vitro study was to evaluate the effect of different
scans on the marginal fit of CAD-CAM monolithic zirconia crowns
fabricated by 3D laboratory scanning and CBCT.
Material and methods
An
extracted maxillary left first molar was prepared and digitized by
using a 3D laboratory scanner (D900; 3Shape) (control group). The tooth
was also scanned by CBCT (i-CAT; Imaging Sciences) to generate a second
virtual 3D model (CBCTscan group). A tooth cast out of polyurethane (PU)
(Zenotec Model; Wieland) was reproduced from the CBCT data by using a
CAD software program (Dental System 2.6; 3Shape) and milling machine
(CORiTEC 550i; imes-icore) and further scanned by using the 3D
laboratory scanner to generate a third virtual 3D model to represent a
clinical scenario where a patient's cast is needed (PU3DLab group). A
monolithic zirconia crown design (cement space: margin 40 μm, 1 mm above
70 μm) was used on the virtual models, and crowns were fabricated out
of presintered zirconia blocks (ZenostarT4; Wieland) by using a 5-axis
milling machine (CORiTEC 550i; imes-icore). The crowns were sintered
(Sinterofen HT-S Speed; Mihm-Vogt), and the vertical marginal
discrepancy (VMD) was measured by ×100-magnification microscopy.
Measurements were made at 384 points in 3 groups of 16 specimens. The
measurements for each specimen were averaged, and VMD mean values were
calculated. The Kruskal-Wallis test was used for the statistical
analysis (α=.05). The Mann-Whitney U test and Bonferroni adjustment were
further used to compare the pairs (α=.017).
Results
The
mean VMD value was 41 μm (median: 38 μm) for the control group, 44 μm
(median: 42 μm) for the CBCTscan, and 60 μm (median: 58 μm) for the
PU3DLab. No significant difference was found between control and
CBCTscan groups (P=.274). However, there was a significant difference between control and PU3DLab and CBCTscan and PU3DLab groups (P<.001).
Conclusions
Marginal
fit of the crowns fabricated by using the 3D laboratory scanner and
through the direct use of CBCT was better than that of the crowns
fabricated by using the workflow that combined the use of CBCT, PU cast,
and 3D laboratory scanner. All tested protocols enabled the fabrication
of monolithic zirconia crowns with a marginal discrepancy smaller than
120 μm.