Low-dose CBCT imaging of alveolar buccal bone adjacent to mandibular anterior teeth— a pilot study


Abstract

Objectives

Accurate description of buccal bone adjacent to mandibular anterior teeth is helpful for planning and monitoring periodontal and orthodontic treatment. Low-dose cone beam computed tomography (LD-CBCT) imaging has shown promising results for very small dental structures in animals. This study asserts that LD-CBCT is sufficiently accurate to measure buccal alveolar bone adjacent to human mandibular anterior teeth.

Materials and methods

Buccal bone level adjacent to 16 mandibular anterior teeth from four human cadavers was measured radiographically using one high-dose (HD) CBCT protocol and two LD-CBCT protocols. The resulting radiographic measurements of buccal bone height (bl) and thickness (bt) were compared with reference probe and reflected-light microscopy measurements. Measurement medians and Bland–Altman plots were calculated, and a linear mixed model was used to compare raters and imaging modalities.

Results

All regression coefficients were approximately 0, indicating high interrater, intrarater, and intermodality agreement. No significant differences were found between reference measurements and CBCT protocols. The mean differences for bl measurements were 0.07 mm (rater 1 [r1]) and 0.12 mm (rater 2 [r2]) for HD-CBCT; 0.07 mm (r1) and 0.13 mm (r2) for LD-CBCT-1; and 0.02 mm (r1) and 0.01 mm (r2) for LD-CBCT-2. For bt measurements, mean differences were 0.02 mm (r1) and 0.02 mm (r2) for HD-CBCT; 0.01 mm (r1) and 0.01 mm (r2) for LD-CBCT-1; and 0.00 mm (r1) and 0.01 mm (r2) for LD-CBCT-2.

Conclusions

Within the limitations of the present study, LD-CBCT seems to be a precise method for describing buccal bone and its thickness adjacent to mandibular anterior teeth in this experimental setting.

Clinical relevance

For the first time, this study showed LD-CBCT produces excellent results and is a reliable modality for imaging buccal bone in vitro. If clinical studies confirm these results, LD-CBCT could enable better treatment planning and monitoring at a radiation dose that is far lower than that of conventional HD-CBCT but similar to that of panoramic views.